Weighted four-parameter logistic fit with back-calculated recovery per calibrator. R² is reported as a response-space diagnostic and is never the acceptance criterion.
Know whether the run was valid — and be able to show how you know.
Four routines evaluate calibration, replicate precision, amplification quality and batch drift. Each records the acceptance parameters it was given, the result it produced, and who ran it. Nothing is assumed on your behalf.
R1 · recorded run · 9-point calibrator series
python measure_quality.py --mutation.Each one refuses to run until it has been told what acceptable means.
Per-marker mean, standard deviation and %CV, with the Grubbs outlier policy applied exactly as configured. Every verdict names the marker and the reason.
Ct calling, amplification efficiency, no-template-control evaluation and ΔΔCt normalisation. Refuses to run until the assay-specific protocol parameters are supplied.
Westgard multirules against the laboratory’s own established control limits, with zone concordance and completeness. The batch being judged never defines the limits used to judge it.
A result the software cannot stand behind is withheld, not estimated.
Most of the engineering is in the refusals. They are listed here because a laboratory evaluating this needs the boundary before the capability.
From a technical review to a controlled pilot, in four steps.
Evaluation copies run against de-identified native exports. No patient data, and no specimen, is required at any point in this pathway.